Reference laser integrated in the device
Reference laser integrated in the device
In order to improve our instrument, allowing us doing spectral analysis without the need of having the reference signal on the target object, we realize a new experimental setup. We have diffused the reference laser in front of the objective and repeated the measure done before in the infrared region.
Results show that is possible to determine the spectral distribution also with this experimental setup. Furthermore, this configuration permits the realization of a more compact instrument.
Schematic layout of the experiment
Test target #1:
Tungsten strip lamp
Result: All the results obtained in the past are confirmed
With this new, more compact, experimental setup it is possible to:
measure the blackbody radiation spectra,
estimate the temperature of a hot body,
do thermal imaging.